- rhk@rubert:~/Ohjelmat/Argyll_V1.0.1/bin$ time ./dispcal -v -d 1 -gs -c 1 -p 0.1,0.5,1 -y l -t 6500 samsung215tw_2
- XRandR 1.2 is faulty - falling back to older extensions
- Setting up the instrument
- Instrument Type: Spyder 2
- Serial Number: 00499106
- Hardwar version: 0x0307
- Place instrument on test window.
- Hit Esc, ^C or Q to give up, any other key to continue:
- Display type is LCD
- Target white = 6500.000000 degrees kelvin Daylight spectrum
- Target white brightness = native brightness
- Target black brightness = native brightness
- Target gamma = sRGB curve
- Display adjustment menu:
- Press 1 .. 7
- 1) Black level (CRT: Offset/Brightness)
- 2) White point (Color temperature, R,G,B, Gain/Contrast)
- 3) White level (CRT: Gain/Contrast, LCD: Brightness/Backlight)
- 4) Black point (R,G,B, Offset/Brightness)
- 5) Check all
- 6) Measure and set ambient for viewing condition adjustment
- 7) Continue on to calibration
- 8) Exit
- Doing check measurements
- Black = XYZ 0.41 0.38 0.52
- Grey = XYZ 45.06 48.44 52.76
- White = XYZ 177.72 190.86 198.21
- 1% = XYZ 3.76 3.91 4.91
- Current Brightness = 190.86
- Target 50% Level = 43.15, Current = 48.44, error = 2.8%
- Target Near Black = 1.91, Current = 4.72, error = 1.5%
- Target white = x 0.3128, y 0.3291, Current = x 0.3136, y 0.3367, error = 5.28 DE
- Target black = x 0.3128, y 0.3291, Current = x 0.2988, y 0.3110, error = 8.23 DE
- Press 1 .. 7
- 1) Black level (CRT: Offset/Brightness)
- 2) White point (Color temperature, R,G,B, Gain/Contrast)
- 3) White level (CRT: Gain/Contrast, LCD: Brightness/Backlight)
- 4) Black point (R,G,B, Offset/Brightness)
- 5) Check all
- 6) Measure and set ambient for viewing condition adjustment
- 7) Continue on to calibration
- 8) Exit
- Commencing device calibration
- patch 6 of 6
- Black = XYZ 0.43 0.38 0.52
- Red = XYZ 88.35 45.91 3.30
- Green = XYZ 59.25 131.02 25.92
- Blue = XYZ 30.58 13.70 169.60
- White = XYZ 177.51 190.36 198.30
- patch 64 of 64
- Initial native brightness target = 190.358819 cd/m^2
- Had to scale brightness from 190.358819 to 180.524263 to fit within gamut,
- corresponding to RGB 0.981837 0.954433 1.000000
- Target white value is XYZ 171.584401 180.524263 196.481137
- Adjusted target black XYZ 0.41 0.38 0.52, Lab 1.89 1.09 -0.84
- Target black after min adjust: XYZ 0.41 0.38 0.52, Lab 1.89 1.09 -0.84
- Gamma curve input offset = 0.026985, output offset = 0.000000, power = 0.000000
- Total Iteration 3, Final Samples = 64 Final Repeat threshold = 0.600000
- Creating initial calibration curves...
- Doing iteration 1 with 16 sample points and repeat threshold of 1.200000 DE
- patch 8 of 16
- real 10m14.920s
- user 0m0.884s
- sys 0m0.056s
- rhk@rubert:~/Ohjelmat/Argyll_V1.0.1/bin$ time ./dispcal -v -d 1 -gs -c 1 -p 0.1,0.5,1 -y l -t 6500 samsung215tw_2
- XRandR 1.2 is faulty - falling back to older extensions
- Setting up the instrument
- Instrument Type: Spyder 2
- Serial Number: 00499106
- Hardwar version: 0x0307
- Place instrument on test window.
- Hit Esc, ^C or Q to give up, any other key to continue:
- Display type is LCD
- Target white = 6500.000000 degrees kelvin Daylight spectrum
- Target white brightness = native brightness
- Target black brightness = native brightness
- Target gamma = sRGB curve
- Display adjustment menu:
- Press 1 .. 7
- 1) Black level (CRT: Offset/Brightness)
- 2) White point (Color temperature, R,G,B, Gain/Contrast)
- 3) White level (CRT: Gain/Contrast, LCD: Brightness/Backlight)
- 4) Black point (R,G,B, Offset/Brightness)
- 5) Check all
- 6) Measure and set ambient for viewing condition adjustment
- 7) Continue on to calibration
- 8) Exit
- Doing check measurements
- Black = XYZ 0.30 0.59 0.30
- Grey = XYZ 45.35 48.98 52.84
- White = XYZ 176.20 189.14 197.03
- 1% = XYZ 3.84 4.31 4.91
- Current Brightness = 189.14
- Target 50% Level = 44.12, Current = 48.98, error = 2.6%
- Target Near Black = 1.89, Current = 4.71, error = 1.5%
- Target white = x 0.3128, y 0.3291, Current = x 0.3133, y 0.3363, error = 5.12 DE
- Target black = x 0.3128, y 0.3291, Current = x 0.2938, y 0.3304, error = 12.16 DE
- Press 1 .. 7
- 1) Black level (CRT: Offset/Brightness)
- 2) White point (Color temperature, R,G,B, Gain/Contrast)
- 3) White level (CRT: Gain/Contrast, LCD: Brightness/Backlight)
- 4) Black point (R,G,B, Offset/Brightness)
- 5) Check all
- 6) Measure and set ambient for viewing condition adjustment
- 7) Continue on to calibration
- 8) Exit
- Commencing device calibration
- patch 6 of 6
- Black = XYZ 0.33 0.60 0.30
- Red = XYZ 87.54 45.69 3.04
- Green = XYZ 58.71 130.17 25.48
- Blue = XYZ 30.31 13.79 168.23
- White = XYZ 176.03 188.97 196.92
- patch 64 of 64
- Initial native brightness target = 188.970132 cd/m^2
- Had to scale brightness from 188.970132 to 179.284450 to fit within gamut,
- corresponding to RGB 0.984339 0.953438 1.000000
- Target white value is XYZ 170.405986 179.284450 195.131734
- Adjusted target black XYZ 0.30 0.59 0.29, Lab 3.00 -6.00 2.82
- Target black after min adjust: XYZ 0.30 0.59 0.29, Lab 3.00 -6.00 2.82
- Gamma curve input offset = 0.042763, output offset = 0.000000, power = 0.000000
- Total Iteration 3, Final Samples = 64 Final Repeat threshold = 0.600000
- Creating initial calibration curves...
- Doing iteration 1 with 16 sample points and repeat threshold of 1.200000 DE
- patch 16 of 16
- Brightness error = 0.316918 cd/m^2
- White point error = 0.957146 deltaE
- Maximum neutral error (@ 0.608811) = 1.004709 deltaE
- Average neutral error = 0.669819 deltaE
- Number of measurements taken = 20
- Computing update to calibration curves...
- Doing iteration 2 with 32 sample points and repeat threshold of 0.848528 DE
- patch 32 of 32
- Brightness error = -0.559196 cd/m^2
- White point error = 0.527274 deltaE
- Maximum neutral error (@ 0.948889) = 0.834208 deltaE
- Average neutral error = 0.491012 deltaE
- Number of measurements taken = 48
- Computing update to calibration curves...
- Doing iteration 3 with 64 sample points and repeat threshold of 0.600000 DE
- patch 64 of 64
- Brightness error = -1.130862 cd/m^2
- White point error = 0.536609 deltaE
- Maximum neutral error (@ 0.000000) = 1.315364 deltaE
- Average neutral error = 0.602301 deltaE
- Failed to meet target 0.600000 delta E, got worst case 0.663169
- Number of measurements taken = 273
- The instrument can be removed from the screen.
- Written calibration file 'samsung215tw_2.cal'
- real 32m24.039s
- user 0m2.120s
- sys 0m0.120s
- rhk@rubert:~/Ohjelmat/Argyll_V1.0.1/bin$ time ./targen -v -d 3 -f 500 samsung215tw_2
- RGB test chart
- Full spread patches = 500
- Multi-dimention steps = 2
- Added fixed 10/11Stats after fixed points: Min = 0.000000, Average = 0.433013, Max = 0.866025
- Percep: Min = 0.000000, Average = 0.433013, Max = 0.866025
- Added 497/497
- Stats after seeding points: Min = 0.000000, Average = 0.090924, Max = 0.123175
- Percep: Min = 0.000000, Average = 0.102167, Max = 0.123175
- Maxmv = 0.054288, stats Min = 0.000000, Average = 0.085182, Max = 0.124155
- Percep: Min = 0.000000, Average = 0.099997, Max = 0.124155
- Maxmv = 0.034997, stats Min = 0.000000, Average = 0.083948, Max = 0.121818
- Percep: Min = 0.000000, Average = 0.099493, Max = 0.121818
- Maxmv = 0.032694, stats Min = 0.000000, Average = 0.083200, Max = 0.119056
- Percep: Min = 0.000000, Average = 0.098903, Max = 0.119056
- Maxmv = 0.036801, stats Min = 0.000000, Average = 0.082798, Max = 0.117528
- Percep: Min = 0.000000, Average = 0.098572, Max = 0.117528
- Maxmv = 0.027209, stats Min = 0.000000, Average = 0.082494, Max = 0.117182
- Percep: Min = 0.000000, Average = 0.098419, Max = 0.117182
- Maxmv = 0.023460, stats Min = 0.000000, Average = 0.082266, Max = 0.117054
- Percep: Min = 0.000000, Average = 0.098388, Max = 0.117054
- Maxmv = 0.026606, stats Min = 0.000000, Average = 0.082080, Max = 0.116617
- Percep: Min = 0.000000, Average = 0.098366, Max = 0.116617
- Maxmv = 0.051792, stats Min = 0.000000, Average = 0.081783, Max = 0.115730
- Percep: Min = 0.000000, Average = 0.098147, Max = 0.115730
- Maxmv = 0.025099, stats Min = 0.000000, Average = 0.081603, Max = 0.117650
- Percep: Min = 0.000000, Average = 0.097999, Max = 0.117650
- Maxmv = 0.022320, stats Min = 0.000000, Average = 0.081518, Max = 0.118084
- Percep: Min = 0.000000, Average = 0.097993, Max = 0.118084
- Maxmv = 0.023235, stats Min = 0.000000, Average = 0.081477, Max = 0.117187
- Percep: Min = 0.000000, Average = 0.098012, Max = 0.117187
- Maxmv = 0.022126, stats Min = 0.000000, Average = 0.081386, Max = 0.117872
- Percep: Min = 0.000000, Average = 0.098029, Max = 0.117872
- Maxmv = 0.019966, stats Min = 0.000000, Average = 0.081312, Max = 0.117173
- Percep: Min = 0.000000, Average = 0.098034, Max = 0.117173
- Maxmv = 0.019301, stats Min = 0.000000, Average = 0.081173, Max = 0.114213
- Percep: Min = 0.000000, Average = 0.098024, Max = 0.114213
- Maxmv = 0.033811, stats Min = 0.000000, Average = 0.081107, Max = 0.115725
- Percep: Min = 0.000000, Average = 0.098006, Max = 0.115725
- Maxmv = 0.021651, stats Min = 0.000000, Average = 0.081013, Max = 0.114040
- Percep: Min = 0.000000, Average = 0.097966, Max = 0.114040
- Maxmv = 0.029977, stats Min = 0.000000, Average = 0.080921, Max = 0.114145
- Percep: Min = 0.000000, Average = 0.097867, Max = 0.114145
- Maxmv = 0.015174, stats Min = 0.000000, Average = 0.080943, Max = 0.114754
- Percep: Min = 0.000000, Average = 0.097986, Max = 0.114754
- Maxmv = 0.015739, stats Min = 0.000000, Average = 0.080974, Max = 0.114599
- Percep: Min = 0.000000, Average = 0.098053, Max = 0.114599
- Maxmv = 0.020790, stats Min = 0.000000, Average = 0.080937, Max = 0.115695
- Percep: Min = 0.000000, Average = 0.098049, Max = 0.115695
- Stats after opt: Min = 0.000000, Average = 0.080937, Max = 0.115695
- Percep: Min = 0.000000, Average = 0.098049, Max = 0.115695
- Total number of patches = 500
- Execution time = 1.120000 seconds
- Execution time = 1.120000 seconds
- real 0m1.191s
- user 0m1.132s
- sys 0m0.004s
- rhk@rubert:~/Ohjelmat/Argyll_V1.0.1/bin$ time ./dispread -v -d 1 -c 1 -y l -p 0.1,0.5,1 -k samsung215tw_2.cal samsung215tw_2
- XRandR 1.2 is faulty - falling back to older extensions
- Number of patches = 500
- Setting up the instrument
- Instrument Type: Spyder 2
- Serial Number: 00499106
- Hardwar version: 0x0307
- Place instrument on test window.
- Hit Esc, ^C or Q to give up, any other key to continue:
- patch 500 of 500
- The instrument can be removed from the screen.
- Written 'samsung215tw_2.ti3'
- real 23m17.987s
- user 0m0.044s
- sys 0m0.108s
- rhk@rubert:~/Ohjelmat/Argyll_V1.0.1/bin$ time ./colprof -v -A "Samsung" -M "215tw" -D "11.07.2008" -C "Risto H. Kurppa" -q m samsung215tw
- rhk@rubert:~/Ohjelmat/Argyll_V1.0.1/bin$ time ./colprof -v -A "Samsung" -M "215tw" -D "11.07.2008" -C "Risto H. Kurppa" -q m samsung215tw_2
- No of test patches = 500
- Estimating white point
- Approximate White point XYZ = 0.946750 0.999950 1.091200, Lab = 99.998066 -3.026365 -19.552099
- Creating optimised per channel curves
- Initial White Point XYZ 0.946750 0.999950 1.091200, Lab 99.998066 -3.026365 -19.552099
- About to optimise temporary matrix
- ...................
- About to optimise input curves and matrix
- .....................................................................................
- About to optimise output curves and matrix
- ....................................
- About to optimise input curves and matrix again
- .
- About to optimise input, matrix and output together
- .................................................
- About to adjust a and b output curves for white point
- About to create grid position input curves
- Create final clut from scattered data
- ***************************************************************************************************************************************************************************************************************************************************************************************************************************************************************************************
- Doing White point fine tune:
- Before fine tune, WP = XYZ 0.963413 0.999432 0.825332, Lab 99.978041 -0.041339 -0.072764
- After fine tune, WP = XYZ 0.946043 0.999409 1.091778, Lab 99.977141 -3.059972 -19.626922
- Creating fast inverse input lookups
- Compensate scattered data for input curves
- White point XYZ = 0.946043 0.999409 1.091778, Lab = 99.977141 -3.059972 -19.626922
- Find black point
- Black point XYZ = 0.001637 0.003438 0.002014, Lab = 3.105457 -6.774116 1.551358
- Display Luminance = 175.965747
- Done A to B table creation
- Setting up B to A table lookup
- Creating B to A tables
- 100%
- Done B to A tables
- profile check complete, peak err = 3.242100, avg err = 0.741296
- real 0m12.380s
- user 0m11.293s
- sys 0m0.260s
- rhk@rubert:~/Ohjelmat/Argyll_V1.0.1/bin$ time xcalib -s 0 -v samsung215tw_2.icc
- vcgt found
- channels: 3
- entry size: 16bits
- entries/channel: 256
- tag size: 1554
- Red Brightness: 0.000000 Contrast: 98.431374 Max: 0.984314 Min: 0.000000
- Green Brightness: 0.000000 Contrast: 95.294121 Max: 0.952941 Min: 0.000000
- Blue Brightness: 0.000000 Contrast: 100.000000 Max: 1.000000 Min: 0.000000
- X-LUT size: 256
- real 0m0.036s
- user 0m0.004s
- sys 0m0.000s
- rhk@rubert:~/Ohjelmat/Argyll_V1.0.1/bin$